The latest generation with innovative technique and future-oriented design.

Basic Equipment:

  • capacitive Multi-Touch-Display in the front
  • defined keys for elementary functions of the installation such as start/stop, fixed-value/automatic operation, interruption, light, etc.
  • USB-Connection to save measuring data on USB Stick
  • Ethernet-Interface / RS232-Interface
  • 2 Digital Out, to switch test specimen ON/OFF
  • adjustable software-temperature-limiter min./max.
  • shelf, stainless steel
  • entry port Ø 50 mm on the right side

Your benefits:

  • low consumption data
  • minimum noise level
  • user-friendly operation and programming
  • uniform operation due to identical Multi-Touch-Display for all chamber sizes
  • service-friendly construction
  • international standards are fulfilled (CE-Conformity)

Test space capacity in litres:

  • 25 / 50

Temperature range:

  • -40 °C / +180 °C
    -65 °C / +180 °C

Design:

Control:
Multi-Touch-Display with integrated program controller
Test space:
stainless steel grade 1.4301
Refrigerants:
environmental friendly refrigerants

Options:

  • add. entry ports Ø 50 mm, 80 mm und 125 mm
  • rack incl. 1 shelf
  • table with height 330 or 515 mm
  • window, multiple glazed
  • add. Pt 100 for temperature measuring on test specimen reversible to controller function
  • air dryer system
  • temperature protection for test specimen
  • CID software for programming and documentation
  • variable air speed in the test space
  • RS232 interface changeable to USB
  • notch

Type T -40/25 T -40/50 T -65/50
Test Space litres 25 50 50
Test Space Dimensions in mm height, appr.
width, appr.
depth, appr.
300
350
240
400
400
320
400
400
320
Overall Dimensions in mm height, appr.
width, appr.
depth, appr.
860
510
580
1030
560
660
1030
560
840
Design bench top cabinet
Temperature Tests
Temperature Range in °C -40/+180 -40/+180 -65/+180
Temp. Change Rate Cooling acc. IEC 60068-3-5 in K/min 6,5 5,0 4,0
Temp. Change Rate Heating acc. IEC 60068-3-5 in K/min 8,0 5,0 6,0
max. Thermal Load at +20°C in W 400 400 400
Temperature Fluctuation in K ≤± 0,3 ≤± 0,3 ≤± 0,3
Nominal Voltage 230 V +6% -10%, 1/N, 50 Hz
Nominal Power in kW appr. 2,2 2,2 2,8
Condenser air cooled
Weight in kg appr. 90 130 160
Noise level in dB(A) 50 50 < 52

All figures are average values which have been obtained at a temperature of +25° C, without test specimens, without thermal load and without options. We reserve the right to make alterations due to technical development.

Other chamber sizes and options on request.